醫療用內窺鏡設備定期保養一般由儀管員配合工程技術人員完成,它是一項不斷循環進行的有組織有計劃的維修措施,有利于出現故障后的查找,定期保養的內容和時間,一般可以分3個等級。
The regular maintenance of medical endoscope equipment is generally completed by the instrument administrator in cooperation with the engineering and technical personnel. It is an organized and planned maintenance measure that is constantly circulating, which is conducive to the search after failure. The content and time of regular maintenance can be generally divided into three levels.
(1)一保:一般可以一個月一個季度進行一次.主要內容除廠日常維護的工作外,可以拆開機殼,各處積塵,污垢、異物、緊固螺絲、添加潤滑劑;檢查務器件、元件有無磨損、變形、燒蝕、擊穿、松動,受潮,老化、接地不良等情況;檢測各組電源電壓及紋波,檢查高壓部件運行和接觸情況等。
(1) First warranty: generally, it can be carried out once a month to a quarter. In addition to the daily maintenance work of the plant, the main content is to dismantle the casing, remove dust, dirt, foreign matters, tighten screws and add lubricants everywhere; check whether the service components and elements are worn, deformed, ablated, broken, loose, damped, aged, poorly grounded, etc; Check the power supply voltage and ripple of each group, and check the operation and contact of high-voltage components.
(2)二保:一般可以半年一年進行一次。主要內容除做好一保外,可以對整機控制臺上的各個儀表及操作控制系統的靈敏度,精度進行測試校正和計量檢定,更換高壓發生器絕緣汕等到期的損耗品,對電路中各測試點的電壓,波形進行系統檢測和做拉偏試驗。
(2) Second guarantee: generally, it can be carried out once every six months to one year. In addition to the first guarantee, it can test, calibrate and measure the sensitivity and accuracy of each instrument and operation control system on the console of the whole machine, replace the expired loss products such as the insulation of the high-voltage generator, and systematically test the voltage and waveform of each test point in the circuit and conduct pull deflection test.
(3)三保:一般可以2-4年進行一次:主要內容除做好二保外,必要寸可以將整機進行全部拆卸子以清洗檢修,超過使用期的元器件應盡量更換或修復,應對儀器進行較為徹底的調試,恢復其工作精度和性能,達到或超過新機的程度是完全有可能的。
(3) Three Guarantees: generally once every 2-4 years: in addition to the two guarantees, the whole machine can be disassembled for cleaning and maintenance if necessary. The components beyond the service life shall be replaced or repaired as much as possible. The instrument shall be fully and thoroughly debugged to restore its working accuracy and performance, and it is completely possible to reach or exceed the level of the new machine.
醫療用內窺鏡設備保養
Medical endoscope equipment maintenance
在醫院各科室中,醫療用內窺鏡的使用極其嚴格,內窺鏡對使用環境的要求:室內環境以暗室環境為主,避免強光直射,手術開始關閉日光燈,用無影燈照明,術前室內的溫度保持在22℃~25℃、濕度50%左右、防酸、防堿、防蒸氣。避免強電、電磁場的干擾。手術中確保電壓穩定,連續穩壓器維持電壓220V以免損壞精密儀器。
In all departments of the hospital, the use of medical endoscopes is extremely strict. The requirements for the use environment of endoscopes are as follows: the indoor environment is mainly dark room environment, avoiding direct light, the operation starts to turn off the fluorescent lamp, lighting with shadowless lamp, the indoor temperature before the operation is kept at 22 ℃ ~ 25 ℃, the humidity is about 50%, acid proof, alkali proof and steam proof. Avoid the interference of strong electricity and electromagnetic field. During the operation, the voltage is stable, and the continuous voltage regulator maintains the voltage of 220V to avoid damaging the precision instrument.
醫用內窺鏡在不同的時期都促進了醫學事業的不斷發展。今后隨著電子技術及其他科學技術的不斷進步,相信其技術會有更廣更深的發展。
Medical endoscopy has promoted the continuous development of medical cause in different periods. In the future, with the continuous progress of electronic technology and other science and technology, I believe that its technology will have a broader and deeper development.
醫用內窺鏡技術發展到今天,已經顯示出它的強大生命力,相信明天會做出更輝煌的貢獻。
With the development of medical endoscopy technology, it has shown its strong vitality. I believe that it will make more brilliant contributions tomorrow.
Composition and diagnosis of endoscopy in hospital http://www.veeker.xyz/ This article is dedicated to the maintenance of endoscope equipment. I believe that you will use relevant knowledge in actual work. If you want to know more, you can pay attention to Jinan Jiangren Medical Equipment Co., Ltd.