內窺鏡攝像系統是內鏡微創手術中不可缺少的設備之一,它可以把反映腔內情況的像傳導出來,通過醫用監視器,方便醫生觀察清晰細節的腔內組織,便于手術醫生更準確的判斷。也可以把外面強光源的光導入器官內,照亮檢查部位。下面內窺鏡維修公司來講講內窺鏡攝像系統圖像偏色處理方法。
Endoscopic camera system is one of the indispensable equipment in endoscopic minimally invasive surgery. It can transmit the image reflecting the situation in the cavity. Through the medical monitor, it is convenient for doctors to observe the clear details of the intracavity tissue and for surgeons to judge more accurately. The light from the external strong light source can also be introduced into the organ to illuminate the inspection site. Next, we will talk about the image bias processing method of the endoscope camera system.
內窺鏡攝像系統圖像偏色處理方法
Image color deviation processing method for endoscope camera system
1、白平衡是否設置正確
1. First, whether the white balance is set correctly
可以重新做白平衡,對干凈紗布對白時,鏡子前端不能有碘伏或其他顏色,圖像窗要顯示全白色,才能保證對白后的顏色正確,對白時間2-3秒,直到提示白平衡完成,然后再次使用,查看顏色是否還原。
The white balance can be done again. When the clean gauze is used for dialogue, the front end of the mirror cannot have iodophor or other colors. The image window must display full white to ensure the correct color after the dialogue. The dialogue lasts 2-3 seconds until the white balance is prompted. Then use it again to see if the color is restored.
2、冷光源對圖像顏色的影響
2. Influence of cold light source on image color
冷光源作為內窺鏡的照明光源的主要指標是亮度和色溫,內鏡末端出來的光不是閃爍的白光,而是帶有顏色的光,判斷冷光源故障,任何一種顏色的缺失將導致圖像偏色。
As the main indicator of the endoscope's lighting source, the cold light source is brightness and color temperature. The light coming out of the end of the endoscope is not flashing white light, but light with color. Judge the failure of the cold light source. The absence of any color will lead to image color deviation.
如果白平衡正確后和排除其他配套設備的影響下都無法還原真實的顏色時,有可能是導線老化,信號傳遞缺陷引起的內窺鏡攝像系統圖像偏色,還有可能是攝像頭電路元件老化變質,又或者是由攝像頭CCD老化引起的圖像偏色,這需要根據不同的原因,替換相應的配件則可。
If the true color cannot be restored after the white balance is correct and the influence of other supporting equipment is excluded, it may be caused by the aging of the wire, the color deviation of the image of the endoscope camera system caused by the signal transmission defect, or the aging of the camera circuit components, or the color deviation of the image caused by the aging of the camera CCD, which requires replacing the corresponding accessories according to different reasons.
內窺鏡設備滿足耳鼻喉科、普外科、泌尿外科、婦科、骨科、神經外科、整形外科等科室常規檢查和手術,當然在使用前后要記得按照相關規范來清潔。更多相關內容事項就來我們網站
http://www.veeker.xyz進行了解吧!
Endoscopic equipment shall meet the requirements of routine inspection and operation in otorhinolaryngology, general surgery, urology, gynecology, orthopaedics, neurosurgery, plastic surgery and other departments. Of course, before and after use, please remember to clean and disinfect according to relevant specifications. Come to our website for more information http://www.veeker.xyz Get to know!