1.手術時嚴禁大力彎折或碰撞;
During surgery, it is strictly prohibited to bend or collide vigorously;
2.使用輔助治設備時要保證在視野內操作,當窺鏡配合激光汽化、高頻電切、微波等光電技術進行手術時,應注意窺鏡前端與點的距離,保證窺鏡前端不被電擊或燒灼;
2. When using auxiliary treatment equipment, it is necessary to ensure safe operation in the field of vision. When performing surgery with optoelectronic technologies such as laser vaporization, high-frequency electrocision, and microwave, attention should be paid to the distance between the front end of the endoscope and the treatment point to ensure that the front end of the endoscope is not electrocuted or burned;
3.術中調整視野角度時需保護好鏡子本身,鏡子與鞘管鏈接時必須對準位置并用卡鎖固定,不易過猛;
3. Protect the mirror itself when adjusting the visual angle during the operation. When the mirror is connected with the sheath, it must be aligned to the position and fixed with a lock, which is not easy to be too violent;
4.建議單獨存放避免與其它器械碰撞,必須輕取輕放以免碰撞鏡子,鏡子傳像系統是玻璃柱體易碎;
4. It is recommended to store separately to avoid collision with other devices. It must be handled with care to avoid collision with the mirror. The mirror image transmission system is a glass column that is fragile;
5.建議手術前用溫水浸泡,涂防霧油;
5. It is recommended to soak in warm water and apply antifogging oil before surgery;
6.在清洗和的時候,
內窺鏡設備維修人員建議把鏡子和別的手術器械分開,用塑料盆清洗;
When cleaning and disinfecting, the endoscope equipment maintenance personnel recommend that the mirror be separated from other surgical instruments and cleaned with a plastic basin;
7.前要把鏡頭清洗干凈,否則異物在后積累在鏡頭上,以影響圖像質量。當鏡頭上異物過多時,可用棉簽涂一些鏡子附帶的清潔劑在鏡面上,然后用清水清洗干凈;
Clean the lens before disinfection, otherwise foreign matter will accumulate on the lens after disinfection, affecting image quality. When there are too many foreign matters on the lens, you can use a cotton swab to apply some of the cleaning agent attached to the mirror on the surface, and then clean it with clean water;
8.盡量不要頻繁更換方法,這樣對鏡子的密封有好處;
8. Try not to frequently change the disinfection and sterilization methods, which is beneficial to the sealing of the mirror;
9.禁止用超聲波清洗鏡子,盡量不要用生理鹽水清洗鏡子,否則容易生銹跡;
9. It is prohibited to use ultrasound to clean the mirror, and try not to use physiological saline to clean the mirror, otherwise it will easily rust;
10.當高溫高壓完鏡子后,一定要讓鏡子自然冷卻,禁止用冷水冷卻;
10. After high-temperature high-pressure disinfection of the mirror, it is necessary to allow the mirror to naturally cool, and it is prohibited to use cold water for cooling;
11.所以的鏡子浸泡時間不能超過60分鐘;
11. The immersion disinfection time of all mirrors should not exceed 60 minutes;
12.高溫高壓:建議盡量不要用高溫高壓,但在緊急時也可以進行高溫高壓。條件:建議控制高溫高壓時間及充分的自然冷卻。
12. High temperature and high pressure disinfection: It is recommended not to use high temperature and high pressure disinfection as much as possible, but it can also be used in emergency situations. Conditions: It is recommended to control the sterilization time of high-temperature and high-pressure disinfection and sufficient natural cooling.
"The precautions for the use, cleaning, and disinfection of endoscope equipment have been explained. If you have any doubts or needs about this, please come to our website." http://www.veeker.xyz Consult and understand!